Kipp & Zonen CGR4 Pyrgeometer
Kipp & Zonen CGR4 Pyrgeometer

Kipp & Zonen CGR4 Pyrgeometer

Product Number: 363900

Overview

CGR4 pyrgeometer is an instrument for the highest quality scientific measurements. The specially designed meniscus dome provides a 180° field of view with negligible directional response error. A hard-carbon coating on the outside of the dome smoothes the spectral response and provides extra protection to the silicon surface. The excellent thermal stability of the dome construction and coupling to the instrument body eliminates the need for dome temperature measurements or dome shading. CGR4 can be fitted with the CVF4 ventilation unit to further improve its performance. A waterproof socket is fitted for the signature yellow signal cable, which is available in a range of lengths pre-wired to the waterproof plug. The integral bubble level is raised to the top of the housing and can be viewed without removing the redesigned snap-on sun shield, which also covers the connector. The connector with gold-plated contacts allows for easy exchange and re-calibration. The screw-in drying cartridge is easy to remove and the replacement desiccant is supplied in convenient refill packets.

Specifications

-
Technical Attributes
Ambient Temperature -40 - +176 °F (-40 - +80 °C)
Analog Output Yes
Analogue Ouput Signal -5 - 5 mV
Analogue Outputs -5 - 5 mV
Cable Connection 8 pin plug
Cable Length 33, 82, or 164 ft (10, 25 or 50 m)
Casing IP Rating IP67
Definiton of Measured Values Long-wave atmospheric radiation
Digital outputs N.A.
Dimensions (free text) Ø 5.9 in (150 mm), H 3 in (76.6 mm)
Field of View 180 °
Housing material Anodised aluminum
Measurement Range EU 4.4 - 50 µm
Non-linearity ˂ 1 %
Non-stability ˂ 1 %
Operating humidity 0 - 100 %
Operating temp range 1 -40 - +176 °F (-40 - +80 °C)
Power Consumption none
Power input none
Response Time ˂ 18 s
Sensitivity 5 - 15 µV/W/m²
Temperature Dependency ˂ 1% (-4 °F - +122 °F)
Temperature range -40 - +176 °F (-40 - +80 °C)
Weight EU 1.3 lbs (0.6 kg)
Zero offset A 4 W/m²
Zero offset B 2 W/m²