Frequently Asked Questions
Technical answers and practical guidance for our measurement instruments and systems including installation, operation and calibration.
Frequently asked questions (FAQs)
Pyranometers
This error is related to the zero offset type A. Normally this zero offset is present when the inner dome has a different temperature from the cold junctions of the sensor. Practically this is always the case when there is a clear sky. Because of the low effective sky temperature (<0 °C) the earth surface emits roughly 100 W/m2 longwave infrared radiation upwards. The outer glass dome of a pyranometer also has this emission and is cooling down several degrees below air temperature (the emissivity of glass for the particular wavelength region is nearly 1). The emitted heat is attracted from the body (by conduction in the dome), from the air (by wind) and from the inner dome (through infrared radiation). The inner dome is cooling down too and will attract heat from the body by conduction and from the sensor by the net infrared radiation. The latter heat flow is opposite to the heat flow from absorbed solar radiation and causes the well-knownl known zero depression at night. This negative zero offset is also present on a clear day, however, hidden in the solar radiation signal.
Zero offset type A can be checked by placing a light and IR reflecting cap over the pyranometer. The response to solar radiation will decay with a time constant (1/e) of 1 s, but the dome temperature will go to equilibrium with a time constant of several minutes. So after half a minute the remaining signal represents mainly zero offset type A.
Good ventilation of domes and body is the solution to reducing zero offsets even further. Kipp & Zonen advises the CVF 3 Ventilattion Unit for optimal ventilation and suppression of zero offset type A. Using the CVF 3 zero offset type A will be less than 3 W/m2.

Radiation incident on a flat horizontal surface originating from a point source with a defined zenith position will have an intensity value proportional to the cosine of the zenith angle of incidence. This is sometimes called the ‘cosinelaw’ or ‘cosine-response’. Ideally a pyranometer has a directional response which is exactly the same as the cosine-law. However, in a pyranometer the directional response is influenced by the quality, dimensions and construction of the domes. The maximum deviation from the ideal cosine-response of the pyranometer is given up to 80° angle of incidence with respect to 1000 W/m2 irradiance at normal incidence (0°).

For CMP 3 < 20 W/m2 and for CMP 22 < 5 W/m2
The CMP series can also be used under water, the depth is limited to 1 meter and can only be used for short measurements.
It is advisable not to keep the Pyranometer of the CMP series under water for longer than 30 minutes.
The SP Lite2 pyranometer and the PQS 1 PAR Quantum Sensor can be used for a longer period under water, the depth is limited to 2 meters. Please also take “breaking of light on the water surface” in consideration.

The 50 % points are the wavelengths where the output of the instrument is 50 % reduced with 100 % input.

|
|
CMP 3 |
CMP 6 |
CMP 11 |
CMP 21 |
CMP 22 |
|
WMO |
Moderate quality |
Good quality |
High quality |
High quality |
High quality |
|
ISO |
Spectrally Flat Class C |
Spectrally Flat Class B |
Spectrally Flat Class A |
Secondary Standard |
Secondary Standard |
There are some exceptions:
- CMP22 has a bandwidth of 200-3600nm (Quartz glass )
- SP Lite has a bandwidth of 400-1100nm (silicon photo diode)
- CMP3 has a bandwidth of 300-2800nm
Simply cover the CMP 11 so it is fully dark (in box with cloth etc.) Log the data over a period that disturbance is expected, at least one day.
If the data is zero no problem is to be expected.
No, we do not have filters for any of our pyranometers. The only way to do this in a correct way is to use a filter dome. Otherwise the directional response would be affected.
The AMPBOX is the best solution. You will need a suitable PSU and a shunt resistor of 500 Ω to convert the current output (4..20mA) to a voltage output of 2-10V , or you will need a shunt resistor of 50 Ω to convert the current to a voltage output of 0.2-1V.

CMP 6 in combination with PQS1 PAR Quantum Sensor is advised. CMP 6 for outside usage to measure Global solar radiation. PQS1 to measure PAR radiation inside which is most sensitive for plants and crops.

For this application the CMP10 and SMP10 are advised as they have an internal drying cartridge that will last for at least 10 years.
Please note that the pyranometer needs to be mounted in the same angle (POA) as the PV panel.
For users that prefer the desiccant visible Kipp & Zonen offers the CMP11 and SMP11 with visible and user changeable desiccant.

Our SP-Lite is based on a silicon diode which has a response from 400 – 1100 nm.
The advantage is the response time, which is as fast as any PV panel ( milli seconds).
The disadvantage is that not all PV panels have the same spectral range.
A thermopile pyranometer covers the full spectral range of the sun and will give a more accurate measurement of the total (global) solar radiation.
The output from thermopile Pyranometers, such as our CMP Series, is very low – typically around 10 milli-volts on a clear sunny day. To resolve changes of 1 W/m2 requires an ADC with an accuracy and resolution of around 5 micro-volts. These PC interfaces are very expensive and difficult to find in a form that is easily interfaced to the PC. This is why meteorological data loggers are normally used that can cope with the low signal levels.
NIST in the USA supplies calibration services to industry – in case of light they characterise sensors, detectors and lamps for use in manufacturing and for luminance measurement (LUX).
They are not set up for the calibration of sensors for solar radiation and they are not a traceable reference.
The only accepted world standards for the calibration of radiometers for the measurement of global or direct broadband solar radiation are as below:
ISO 9059 Calibration of Field Pyrheliometers by Comparison to a Reference Pyrheliometer
ISO 9060 Specification and Classification of Instruments for Measuring Hemispherical Solar and Direct Solar Radiation
ISO 9846 Calibration of a Pyranometer Using a Pyrheliometer Guide to Meteorological Instruments and Methods of Observation, Fifth ed., WMO-No. 8
The Zero Offset A can also be reduced by using a Ventilation Unit CVF 3.
The CMP 21 is individually tested and the temperature compensation is optimised. It is ±1% from -20 to +50°C. However, from -10 to +40°C it is within ± 0.5%, typically ± 0.3%. In addition a temperature sensor is fitted and the temperature response curve is supplied. Each CMP 21 has the directional (cosine) response tested, and this is also supplied. This means that for the serious scientist the irradiance values can be corrected for temperature and solar elevation – increasing the accuracy. This is not possible with the CMP 11.
BSRN requirements state that the solar radiometers must be fitted with an internal temperature sensor and the data recorded, so CMP 21 is compliant to this, but CMP 11 is not.
The digital RS-485 output can operate over cable lengths up to 1000 m, depending on the baud-rate used.
Pyrgeometers
Net Radiometers
Standard = extended
White = white
Green = blue
Black = black
Because the Pt-100 (unlike a thermocouple) needs current, it is advised to keep this current as low as possible to avoid self-heating of the Pt-100 by its own current. The Pt-100 measuring device (like our data loggers CC 48, CR10X) has a fixed current, in such a way that the voltage over the Pt-100 is matched with the Pt-100 (voltage) measuring input of these loggers.
In general the current for a Pt-100 is indeed between 0.1 and 1 mA. This would result (@ 0ºC) in a voltage over the Pt-100 of 10 mV or 100 mV. Therefore the current can also be selected depending on the available input range of the measuring device. The error introduced by self-heating, when using a 1 mA current, is quite low (< 0.2ºC) also because the Pt-100 is very well connected to the body of the CNR1. When the heater of the CNR1 is on, the error introduced by the heater in measuring the body temperature is typical 2ºC (see manual).
The benefit of a larger current (1 mA) is that electrical disturbances have less effect when the current is larger.
To summarize these facts I would say, 1 mA measuring current is accurate enough, but the output voltage in this case (0.1 Volt) has to match the measuring input range.
The CNR 2 uses two glass domes to cover the pyranometer and two silicon domes to cover the pyrgeometers. It uses TWO thermopile detectors (1 for each of the two pyranometers and 1 for each of the two pyrgeometers) and provides two separate outputs. One NETTO for short wave (solar spectrum) and one NETTO for long wave radiation.(Far Infrared spectrum).
So yes, the CNR 2 has separate thermopiles to measure Far Infrared and Solar radiation and so do the other CNR net radiometers.
The detector from the NR Lite(2) is not protected and I sin direct contact with the weather conditions. Therefore it cools down a lot faster by the wind, which effects the accuracy of the measurements. The NR Lite(2) uses NO dome. It uses only TWO detectors with a PTFE coating and provides ONE single output for NETTO short wave- and long wave radiation. It uses one thermopile to measure the full spectrum of Far Infrared and solar radiation.
CNR 2 uses glass domes for the pyranometers (that measure short wave radiation) that have a bandwidth of 300 nm to 2800 nm. It uses silicon domes for the pyrgeometers (that measure long wave radiation) that have a bandwidth of 4500 nm to 42000 nm. This leaves a gap between 2800 nm and 4500 nm. This is the so called atmospheric window where very little radiation comes in (see picture below).

The NR Lite(2) uses NO domes. It uses two detectors with a PTFE coating which have a bandwidth of 200 nm to 100.000 nm.
UV Radiometers
The bandwidth of the CUV 4 is 280 - 400nm. The 50 % points are defined to be 290 - 385 nm.

This is required because these two factor strongly influence the UV measurement. With the UVIATOR software you can collect the Ozone column data from the OMI satellite data per date, time and location. With this Ozone data and the solar zenith angle the UVIATOR will calculate the optimal correction for every data point.
Yes, the UVS-AB has two independent detection systems.
The UVS-AB has two continuous, simultaneous, analog outputs; one for UV-A and one for UV-B.
See below drawing of the UVS-AB for details.
Beneath the dome and diffuser two sets of filters and detectors are positioned. Detector 1 is 4 times bigger than detector 2. Detector 2 is located exactly in the middle, on top of detector 1. Filter 1 has an opening in the middle for filter 2 plus detector 2. In this way we can make sure both detectors have a 180 degree field of view.

Sun Trackers
Part number for the 24V heater kit is: 12136346.
This kit contains two 50 Watt heaters.
So the current for these heaters is 4.2 A (at 24V), fuse is 5 A slow blow.
If we add up the total power we have:
5A (heaters) + 3.15 (2AP)= 8.15 A
Normal conditions:
4.2 A (heaters) + 1.25A (2AP) = 5.5 A normal
Therefore a 5 Amp power supply will not survive very long.
We recommend to take a power supply that can deliver the 8.15 A. (when heaters are used)
Advised is a shielded cable, where the shield can be the ground connection.
On the 2AP side the wires are connected to the communication board.
On the PC side a 9 or 25 pin Sub D connector is used for the serial port.
This cable length can be 30 meter. We can supply this cable, but it can be bought "around the corner".
The communication software is included with the 2AP.
Standard temp range: 0 - 50 degrees Celsius
With cold cover: -20 - 50 degrees Celsius
With cold cover and heater -50 - 50 Degrees Celsius
Normally the heater is built in, in the factory. However we can supply you with a kit plus instructions to do it yourself.
The 2AP has the following errors:
Time
Setup (leveling)
Calculation (algorithm according to Michailsky error max. is 0.025 degree)
Mechanical (BD = 0.09 and GD =0.045 degree max.)
The first 2 are user controlled and the last 2 are fixed.
Assuming that the leveling is done optimal the only error remaining that can be corrected is time. If we assume that the clock is reset every 1,000,000 seconds (11.5 days) there will be an expected error of 5 seconds. If we assume that the sun rotates 360 degrees in 24 hour then RMS 0.72 * 5 seconds = an actual time error of 3.6 seconds. For a period of 11.5 days the total time error contribution is 0.015 degrees (the diameter of the solar disk is 0.25 degrees). Per year this would result in 0.5 degree.
If this time correction and the check on leveling are done on a regular interval there is no need for a sun sensor. If however this interval can or will not made, the sun sensor will correct for both (leveling and time).
The sun sensor is normally used for first checking the tilt error. Assuming that there is sunshine for at least 2 full days over the full day. This information is stored in an internal log file and used to correct (in combination with PC). Over this period the user has to correct time. (if more than 2 weeks). After this initial run and correction for tilt, the sun sensor is used for time correction. This means that optimal accuracy is maintained without user (time) correction. This means that the 2AP is within specs the whole year without intervention.
Please note that the sensors used on the 2AP also need maintenance on a regular base (drying cartridge and dirt on domes).
Better than giving an error in percentage we would like to show the benefit of not having to correct to clock. The 2AP error in degrees can be calculated as percentage of 360 degrees, but the error in sensor reading depends on the type of sensor.
We enter a room temperature correction, which compensates for initial calibration of the oscillator module. The oscillator module can drift up to ±11 ppm over the 0 °C to 70 °C temperature range. The module can also drift up to ±2 ppm in a year.
The temperature drift is different for each oscillator so cannot be compensated for in firmware.
If you send the command “CO”, the 2AP will cold start. This means ignore all present settings and start without using any previous (possibly wrong) settings.
If you then start the sun tracker program, it will start up with the message “”recovering from cold start””
Then longitude and latitude etc. will be recovered from the .ini file, the time and date of course cannot be stored and has to be entered again. If no further error message is given, the 2AP is most likely operational again.
This could solve the problem, if not please contact us and we will discuss further options.
The best solution is to sell/quote a 2AP Gear Drive with 24VDC, as this would eliminate the altitude restrictions associated with high voltage AC.
The SOLYS Gear Drive can easily handle more, but for mounting more instruments (e.g. 8 pyrheliometers) a larger mounting plate is required.
Yes, like on the SOLYS 2 a special mounting clamp is available for the PMO-6.
When used in cold climates, the heater switches on to keep the interior above -20°C
This is switched automatically and only used when powered from AC.
The cold cover can be used to reduce the required heating power.
This is normally accurate enough, but does not correct for any misalignment or unstable mounting.
The SOLYS has in addition to the radiometers a paint coating to further protect it.
RaZON+
2 year warranty, 5 years on the sensors
Smart Pyranometer(PR1) and Pyrheliometer (PH1) plus cables
Power connector for 24 VDC
Only 1 pyranometer, the pyranometer PR1 or any SMP
On a sunny day, point the system to roughly East and start. The software takes care of fine adjustment and via two small holes in the pyrheliometers you simply can check the final position
If any azimuth adjusting is needed RaZON+ can be rotated on its tripod collar or fine adjustments can be made via WiFi and your Smart device
Data logging is done via the Ethernet or RS-485 port.
- Through Modbus (pull)
- Through ASCII string (push)
- Through Ethernet (database query)
- Through WiFi (database query)
Scintillometers
Plus two 12 VDC power supplies (CVP1 LAS MkII power supply x2), and for mounting, two tripods (heavy duty tripod package x1) or secure mounting structure, if required.
Meteorological sensor kit
Plus two 12 VDC power supplies (CVP1 LAS MkII power supply x2), and for mounting, two tripods (heavy duty tripod package x1) or secure mounting structure, if required.
Validating satellite data / ground truth
Weather forecasting
Irrigation / water management (shortage) - evaporation from rivers, crops and water storage
Hydrology
Micro-meteorology / turbulence studies
Land-atmosphere exchange / boundary layer meteorology
Agriculture, forestry, Biology, plant evapotranspiration, plant interactions and geosciences
Forest fire warning
Optical propagation conditions
Turbulence, including defense and laser propagation
Low maintenance, no moving parts
Low power consumption
Integral data logging
Remote measurement:
Path-averaged Cn2 measurements up to 4.5km
Representative for large area
Comparable to grid box size of numerical models and pixel size of satellite images
No structure / tower influence on measurements (spatial weighting function / no flow distortion by instrument). A tower, building or valley sides can be used to gain height, and they (the support structure) has no or little effect on the measurements as the LAS responds more strongly to the middle of the path and not at all at the ends.
Easy installation - Can measure over terrain which is difficult to access, or which you do not want to disturb.
Does not disturb the measurements and measurement area (such as a protected wildlife area).
Rapid measurements:
Allows study of fast processes, such as plant transpiration and canopy resistance, which change on a timescale of a minute with changing solar radiation and clouds
Point measurements (alternative methods to LAS)
Eddy-Covariance method (need averaging of at least 30 minutes to catch all the eddy sizes. Measures the inner scale of turbulence i.e. the momentum flux, so can be combined with gas flux measurements easily)
Bowen-Ratio Energy Balance method
Flux-profile method (MOST)
Results extremely localised
Influence of structure / flow distortion
The advantages of the LAS MkII are the following:
Low power consumption due to the use of single LED and good collimation of the beam so little light is thrown away. This makes it practical for use with solar panel / battery electrical supply in field applications.
Contains an integrated lens heater to avoid freezing of the instrument or condensation on the window, and provides good data in cold environments.
More compact design with integrated datalogger, display and setup buttons which allows the configuration of the LAS without the need of an additional computer and cables / power supplies, and nothing needs to be taken apart. This makes transport and installation much easier.
Direct connection of the meteorological sensor kit to the receiver instead of to a separate interface or data logger.
Operates over the full data range of Cn2 of 1x10-17 to 1x10-11 (= 6 orders of magnitude).
The transmitter (as well as the receiver) can be tilted to align for optimum signal gain, and minimal power usage.
A supplied GPS antenna, mounted on the receiver, results in very accurate time from the satellites of the Global Positioning System, logged with the measured data.
The complete LAS MkII system is shipped in one rugged aluminium transport case, which can be used in the field to store equipment if needed as the case is waterproof. Lined with custom fitted foam, it protects the LAS instrument from repeated shocks of 10 to 20g without losing calibration or alignment.
A certificate is supplied with each instrument detailing a calibration against a reference instrument, performed outdoors for 2 weeks at the factory.
A large aperture scintillometer (LAS) has less saturation problems compared to a laser scintillometer, or one with a smaller beam diameter, and so can be used over longer distances.
See appendix H of the latest manual, for details of the conversion of the analogue voltage output, UCn2. Note, the manual is written for the improved LAS MkII with GPS, so there will be differences in the hardware, and there are a couple of data formatting issues that are different (voltage ranges).
Most likely the format of your UCn2 data is incorrect for the latest version of Evation, as the software was developed to work with the LAS MkII, which has a different analogue voltage output (positive instead of negative).
For the LAS MkI, the Cn2 voltage output, UCn2, is output as -5 to 0V, where:
-5V analogue voltage output is equivalent to a Cn2 of 1 x 10^-17 [m-2/3]
0V is equivalent to 1 x 10^-12 (m^-2/3)
Cn2 (m^-2/3) = 10^(-12+ UCn2) (this applies to the LAS150, not to the LAS MkII !)
Over non-flat terrain it is a bit more complicated, because now we also need to consider the path-weighting function of the LAS. This weighting function reveals that the centre of the LAS path contributes more to the measured Cn2 data, than near the transmitter and receiver units. This calculation is easily done by using the effective height calculator built-in to the Evation software, which takes care of the weighting function.
Note: for very long path lengths (> 5km), such as when using the XLAS it is also important to consider the earth’s curvature (decreases the height in the centre of the path by approx. 2 m over a path length of 10 km.
More detailed information of deriving the effective height of scintillometers over complex terrain and the effect of atmospheric stability on the effective height can be found in: Hartogensis et al., Derivation of an Effective Height for Scintillometers: La Poza Experiment in Northwest Mexico, Journal of Hydrometeorology, 2003.
The effective height or height of the LAS beam should be measured to 1 cm (a tape measure can be used for this).
See the following publication for more information: van Dinther, D., O. K. Hartogensis, and A. F. Moene, 2013: Crosswinds from a single-aperture scintillometer using spectral techniques. J. Atmos. Oceanic Technol., 30, 3–21.
Figure 1: Theoretical spectrum of a 0.15 m LAS (path length = 1 km, wind speed = 1.5 m/s).
Any type of fluctuations, e.g. caused by tower vibrations that lie within this bandwidth, in particular the ones that lie between 0.5 to 10 Hz can have significant effects on the measurements. It is therefore, strongly recommended to use stable and robust mounting platforms for the LAS units.
The LAS transmitter and receiver both have multiple analogue output signals, which can be measured by most standard data loggers. These signals allow the user to monitor the internal temperature as well as some raw signals to check the performance of the electronics. In most experiments these signals don’t have to be measured.
For general flux measurements two signals are important: the Cn2 signal and Demod signal. Both signals are measured at the receiver unit. The first signal, re-scaled Cn2, provides information of the turbulent intensity of the atmosphere and is used to derive the sensible heat flux (H). It’s range lies between 0 to 2.4 V for the LAS MkII (-5 and 0 Volt for the LAS150). The second signal: the demod signal is a measure of the signal strength and it’s range lies between 0 to 2 V for the LAS MkII (-2 and 0 Volt for the LAS150). The more positive (more negative for the LAS150), the more signal the receiver has. In general the signal strength depends on the distance between the transmitter and receiver and the opacity of the atmosphere.
The reason it is advised to measure the demod signal is that it can help with the interpretation of the Cn2 signal. In some cases the Cn2 can be difficult to understand, e.g. during rainy and foggy periods, while the demod signal shows clearly whether or not the receiver has a signal, or some signal is lost due to the weather.
The surface roughness can also be determined experimentally, using e.g. eddy-covariance stations or from wind profile measurements.
But there are several reasons why LAS measurements of open water is rather complicated:
In general the sensible heat flux (H) is small over lakes compared to the evaporation. As a result the derivation of CT2 from Cn2 becomes sensitive to fluctuations that are the produced by turbulent humidity fluctuations instead of temperature fluctuations. This correction can be expressed as a function of the Bowen-ratio ( ). The smaller the Bowen-ratio the larger the correction. Over water bodies evaporation is (in most cases) dominant over H, resulting in small Bowen-ratio values. It is advised to have accurate Bowen-ratio data.
Typical diurnal course of H over land shows positive values during the day and (small) negative values at night. In this way it is relative easy to select either the unstable or stable solution when processing fluxes (in the Evation software) as the LAS itself cannot see the sign of the flux (and thus stability). Properties of water, such as the ability to store heat (i.e. heat capacity) are very different from land. As a result it is more complicated to predict the diurnal (seasonal) cycle of H. Instead it is advised to measure the gradient of temperature over water in order to determine the sign of H.
For the derivation of the sensible heat flux from CT2, the wind speed and surface roughness are required. As the surface roughness of open water bodies is dependant of the wave height (and thus wind speed) the standard applied flux-profile relationships cannot be used. Instead air-sea relationships have to be considered.
To derive the evaporation from the LAS measurements (i.e. H) the soil heat flux term has to be known, i.e. the amount of heat stored in the ground or in this case stored in the water (G). For land surfaces so-called heat flux plates can be used to measure G. For water bodies this term is very difficult to determine.
In-case one has the ability to check the data in real-time, the demod signal can help to monitor the alignment. A slow decreasing trend of the demod signal can suggest possible changes in the optical alignment (ignoring degradation of the LED).
If there are animals in the area protect the equipment from being pushed over, disturbed, or the cables chewed, by surrounding the installation with electric fencing.
Brewer Spectrophotometer
The drive mechanism can be cleaned by using a clean lint-free cloth with alcohol or with “garage” soap (soap with grains of sand in it). Switch the tracker power off! Remove the rear tracker cover (the cover on the side opposite to the power switch). Use the cloth with alcohol or soap to rub the dirt off the drive plate and the drive shaft. Rotate the tracker to clean the entire drive plate. Be careful not to break the wire of the safety switch. After cleaning the entire drive plate and shaft, rub them once more with a dry piece of clean lint-free cloth to remove any remaining residue of soap/alcohol.
When this is done, rotate the tracker to aim the Brewer approximately at the sun . Put the tracker cover back on and switch on the tracker power. Now the Brewer needs to perform some tracker resets. In the Brewer software, type “PD AZ SR 10” to perform these resets. Watch the data to check that the tracker resets without discrepancies and then put the Brewer software back into its normal schedule.
During the factory testing of the Brewer it undergoes a test in the temperature chamber from 0°C to +45 °C. Standard Lamp measurements are taken throughout this entire temperature range. This is a simulated Ozone measurement based on the halogen lamp inside the Brewer. Although the intensity of the lamp does change with temperature, the wavelength shift is negligible.
After the temperature test, the data of the SL measurements is analysed. During the analysis, the temperature correction coefficients are created. These coefficients compensate for the change in spectral response of the Brewer at the Ozone wavelengths. With the coefficients installed in the Brewer software, the Ozone measurements will not be affected by the temperature of the instrument.
The first step in troubleshooting is doing a full reset (RE command) in the Brewer software. Then try to perform the HG test again.
If the HG test still returns the error message one should find out if this also occurs for tests with the standard lamp. Type SL<enter> in the Brewer software.
If Both the HG test and the SL test fail, then either a motor is not moving correctly or there is a problem with the PMT/photon counting circuitry. Use the maintenance manual for further troubleshooting.
If the SL works but the HG fails, then there might be a problem with your lamp. For single board Brewers: Type AP to get the voltage of the Mercury Lamp (HG lamp). The voltage should be around 10 V. If the Voltage is off by 2 Volts, one should inspect the lamp.
The HG or mercury lamp is the lowest lamp in the lamp housing. Usually, if the lamp needs replacement, the glass will have black spots or the filament will be broken.
If the lamp needs to be replaced, do not touch the quartz envelope with your hands. Use a tissue or a piece of cloth. The lamp should be tightened firmly. Also, from the top, both filaments should be visible.
Some Brewer users prefer to have their Brewers calibrated every year. By doing this, they assure their Brewer data is of the highest quality. Drifts in the instrument are corrected sooner and the regular check with a reference Brewer increases the reliability of the data.
If you would like to discuss calibration of your Brewer at the factory or at your location please contact us.
DustIQ
5MW 1 instrument
40MW 2 instruments
100MW 3 instruments
200MW 4 instruments
300MW 5 instruments
500MW 6 instruments
The DustIQ is installed next to or in between the actual PV modules.


Horticultural Sensors
If the metoffice gives such figures for your area it should be of importance to know the relation between PAR intensity and total global irradiance for at least a solar spectrum at air mass 1.5 (solar elevation 53°).
This relation is for:
Clear sky: 681 W/m² total, 308 W/m² in the 400 to700 nm band 1408 µmol/s.m² PAR
Light cloud cover: 200 W/m² total, 109 W/m² in the 400 to 700 nm band 493 µmol/s.m² PAR
Be aware that many green leaves are highly reflective for near IR but absorb strongly in the 400 to 700 nm band (PAR region). For evaporation only W/m² counts and for photosynthesis only photons counts.
There is the non- stability, non linearity, temperature dependence and the quantum response. Measurements under 1 µmol/m²s are not to be trusted as accurate. (effective limit)
Data Loggers
|
Date |
Time |
A1 |
A2 |
A3 |
|
5-2-2009 |
10:59:03 |
3,079 |
38,763 |
0,693 |
|
5-2-2009 |
11:00:03 |
3,108 |
38,763 |
0,697 |
The picture shows how a PT100 is connected. PT100 uses four input channels nr 1,2 ,3 and 4. Leaving 4 channels left, nr 5,6,7, and 8, to connect other sensors.
Please note that the picture uses 2 analogue inputs for the instrument (differential). Of course you can decide to use it single ended, by connecting the – from the instrument to the GND.

PWR OUT will provide whatever power is connected as power source for the LOGBOX SD.
For example we use a logger with 12 bits resolution and 5 Volt input range.
This means that the 5 Volt is divided in 2¹² steps. This is 5/4096 = 0.00122 Volt per step. So the smallest detectable change on the input is 1.22 mV.
If we connect an SP Lite(2) with a sensitivity of 75 microVolt/W/m^2, one Watt (change) will give 0.075 mV (change) on the output. So we need a change of 1.22 / 0.075 = 16.2 Watt/m² before we see a change in the Logger output.
As you can understand this is not acceptable. A minimum of 1 W/m² should be detected (preferably a factor 10 times better). This can be achieved by lowering the input range to 100 mV (50 times better). Or by selecting a higher resolution e.g. 16 bits (in stead of 12 bits) this is 16 times better ( 2¹⁶- 2¹² = 2⁴ = 16 ).
For example:
Radiation in the morning 4 hours 400 W/m².
Radiation in the afternoon 4 hours 600 W/m².
If you would take a 60 minute interval on the same day:
Morning = 4 intervals of 400 W/m2
Afternoon = 4 intervals of 600 W/m2
Total over the day 4 x 400 + 4 x 600 = 1600 + 2400 = 4000 W/m2 (over hourly interval)
This is 4000 W/m2 per hour or 4000 Wh
The stored interval in METEON is 30 minutes.
During the morning you will get 8 intervals of 400 W/m2
During the afternoon you will get 8 intervals of 600 W/m2
16 intervals (of 30 minutes) with a total of 8 x 400 + 8 x 600 = 8000 W/m2 (over 30 min interval)
This is 8000/2 = 4000 W/m2 per hour or 4000 Wh
Snow Depth Sensors
The SHM31 is mainly designed to measure the snow depth. The measuring principle is not applicable to measure pure ice or water due to strong reflections or translucency.
The power consumption at 12 VDC is roughly: 2.5 W if only the window heater is enabled and 18 W with window and main heater switched on.
Smart Weather Sensors
After a reset (switching On/Off the sensor), every WSwill start up for the first five seconds in UMB-protocol and the standard baud rate of 19200Bd, independent to the configured protocol and baud rate settings of the sensor.
In the first five seconds, you will be able to connect via UMB-config-Tool to the WS. After the first command is sent, the UMB-protocol will stay for ten minutes before it switches back to the original configured protocol. This time will always be extended to ten minutes if a new UMB command was sent.
The wiring plan has changed compared to the wiring plan of older WS device versions.
The new wiring plan is:
+VDD = brown
GND = white
SDI12_Signal = yellow
SDI12_GND = white
This is also mentioned as a note in the latest WS manual in the section 8 Connections.
Inside the smart sensor, a drop size distribution matrix enables to calculate intensity of precipitation and to determine type of precipitation according to fundamental meteorological relationships (Gunn-Kinzer and Hobbs and Locatelli).
Accessories
standard, then the gain is 1 mV = 1 mA
adjusted, then the gain is set to 0 – 1600 W/m2 = 4 – 20 mA
The AMPBOX has a 20 bits A/D on the input and a 16 bits A/D on the output.
The maximum error from the AMPBOX is ±0.05% of span or ±10 µV (over the full temp range)
This means in daily use that the additional error of the AMPBOX is far below 1 W/m2 for all radiometers.
Gain adjust --> 0.1 to 4 mA/mV
Input voltage range --> -12 to 50 mV
As mentioned above the amplification can be adjusted to 0- 1600 W/m2 = 4 – 20 mA.
In this case 100 W/m2 change on the input gives 1 mA change on the output.
The benefit is that two different sensors with different sensitivities have an identical output from the AMPBOX.
The pyrgeometers are adjusted for -300 to +100 W/m2 = 4 to 20 mA.
So 100 W/m2 change on the input gives 4 mA change on the output. The zero point for the pyrgeometer is set to 16 mA. This is because a pyrgeometer can be more negative than positive.
Yes you can, by using a shunt resistor and a suitable power supply unit (PSU). You will need a shunt resistor of 500 Ω to convert the output current (4..20mA) to a voltage output of 2-10V. Or you will need a shunt resistor of 50 Ω to convert the current to a voltage output of 0.2-1V.

For CMP 3 with 16.51 μV sensitivity, this is 0 - 16.51 * 1600 μV =0 - 26.416 mV for 4 - 20mA (delta = 16 mA)
Sensitivity AMPBOX = 26.416 mV per 16 mA = 1.651 mV per 1 mA or 1651 μV per 1 mA
Now connected PAR with 5.59 μV/μmol. 1651 / 5.59 = 295.35 μmol per mA
Therefore:
0 μmol = 4 mA
295.35 μmol = 5 mA etc.
4725.6 μmol = 20mA as maximum input
With a 10 Ohm resistor the 20 mA output from the AMPBOX can be converted to 0.2 V max signal (max for the METEON). But the METEON cannot deal with the zero offset from the AMPBOX (4 mA).